Question
I wonder if you have a multimode (water and methanol/acetonitrile) SEC column that has the lowest MW exclusion limit? The GS220 gave me good performance, but it might improve even more if the exclusion size limit is further lower, e.g. 2 kDa.

I can do custom-order if this is not in your product line.

 
 

Answer
The GS-220 HQ is the smallest pore-size line for our multimode SEC column.
Are you trying to improve the separation of small compounds?
In general, SEC separation can be improved by using multiple columns in series.
Have you thought about using two columns in series?
Do you mind sharing the chromatograms with me, so that I can understand better what type of improvement you are looking for?

Also, what percentage of organic solvents do you need to use to dissolve your sample?
If you do not require multimode/solvent type, we do have smaller pore-sized GFC/GPC columns (e.g. SB-802, which cannot be used with organic solvents).
Another option I could think of is to use small pore-sized RP column, namely ODP2.
This is a polyhydroxymethacrylate based column with average pore size of 40 A. Please note that we use maximum pore size as an indicator for our SEC columns. I do not have the information on maximum pore size for ODP2 in my hands now, but this will have smaller pore size than GS-220. The column can be used with multi-solvent. I hope this will give you some idea of available options and hopefully we can come-up with something suitable for your research.

 
 