Question
how to combine SEC with different pore sizes?

 
 

Answer
The order of the connection of GPC column in series is as following.
smaller
Pump-> Injector -> KF-806-> KF-803-> KF-801-> Detector

That is, the column packed with the larger particle gel is connected at the upper flow line , and the column packed the smaller particle one is at the lower flow line.

Actually, the result of the analysis is same, even if the order of the connection is different.

This is because the tolerance of the pressure or the toughness of the column.
As you may know, the column is easy to be affected by the injection.
And the part of the column which is easy to be affected is the inlet of the column.

Basically, the pressure of the column packed with the larger particle gel is lower than that of one of the column packed with the smaller particle gel.
So, we always use the order mentioned above.

 
 