Skip to main content
Print

Sertraline-Technical question

Record: 174


Category: Column Selection
Product:
Sample: sertraline


Question
sertraline analysis

 
 

Answer
Depends on what they are separating sertraline from, but I think RP will work.
…Could be by ODP50 at high pH condition.
e.g.
http://www.shodex.com/japanese/dc040234.html

 
 
Additional Comments
We are sorry but we do not have the equivalent column to Daicel AD-H column.
And we do not have the separation data for the analysis of Sertraline.

Daicel AD-H column is for chiral separation with using hexane mainly as the eluent.
The column such like this one is often presented as a normal phase chiral separation column.

On the other hand, Shodex CD column cannot be used with such the eluent condition.
The eluent condition for this column is the mixture of certain aqueous solution and some organic solvent.
This is a reversed phase chiral separation column.

However, a reversed phase chiral separation column also seems to work for this analysis.
I will attach a paper regarding the separation of Sertraline with using RP-chiral separation, though the column is not Shodex column.

(See attached file: Sertraline.pdf)

The column used in this paper has beta-cyclodextrine as its ligand. So I think CDBS-453 may work with the similar analysis condition.

image_pdfDownload or Print PDF