SH Column Issue

Record: 405


Category: Column Problem
Product: SH-G
Sample:


Question
using SH column. standards, peak split, she thinks it’s caused by guard column. any cleaning procedure?

 
 

Answer
Please test analytical column itself. Then, if guard is the cause, reverse the flow direction, clean with water at 60 deg C and 0.2-0.3 mL/min.

 
 
Additional Comments