Skip to main content
Print

SH columns and organic solvent

Record: 556


Category: Specifications
Product: SH1011
Sample:


Question
Use of organic solvent is not recommended for SH type columns.
You may try following regeneration steps (pg15-16 on the catalog) but it cannot guarantee to restore the column condition.
-Introduce 0.005N H2SO4 at 50oC using 0.5 mL/min flow rate.
http://shodex.net/data/downloads/SUGAR_series__J_E_No.5092-2.pdf

 
 

Answer
Use of organic solvent is not recommended for SH type columns.
You may try following regeneration steps (pg15-16 on the catalog) but it cannot guarantee to restore the column condition.
-Introduce 0.005N H2SO4 at 50oC using 0.5 mL/min flow rate.
http://shodex.net/data/downloads/SUGAR_series__J_E_No.5092-2.pdf

 
 
Additional Comments

image_pdfDownload or Print PDF