Question
A little bit background info here for your further assistance. The ingredient of our product is showing as below, the lactate is the component we are interested in to qualify. Our plant is using a reserve phase HPLC method+ UV detection with a C18 column. But the baseline of the chromatograms are not flat and the column lifetime is short. The phenomenon is probably due to the icodextrin, which is a starch polymer.
It happens that our Japanese colleagues are using a cation exchange chromatography method with Shim-Pack columns for lactate detection in the same product. So we are searching for the alternative method to resolve our issue in US.
One question here. Can you kindly educate me more about the chemistry behind the specialty of sulfonated polystyrene packing in Shim-Pack SCR-101H and Guard for organic acid detection?

 
 

Answer
As you mentioned, I think SH column will able to separate lactic acid (lactate) and icodextrin.
Since the customer does not need to determine the icodextrin MW in this case (is this correct?) I would recommend SH1011 – so all icodextrin will be eluted as a void; anything bigger than 1K (exclusion limit of SH1011) will not be retained.
From my quick research, icodextrin has MW of 13K-19KDa.
http://shodex.net/index.php?seitenid=1&applic=1085
http://shodex.net/index.php?seitenid=1&applic=1086

Please make sure if icodextrin is 100% soluble in H2SO4 mobile or not.

 
 