Shodex Asahipak NH2P-40 3E

Record: 768


Category: Specifications
Product: NH2P
Sample:


Question
We recently purchased the Asahipak NH2P-40 3E analytical column and I was wondering if there was any recommendation for maximum injection volume and maximum sample load for monosaccharides.

 
 

Answer
Thank you for using the NH2P-40 3E column.
Although below is data for our NH2P-50 4E (4.6 mm ID column), I would recommend using the similar or slightly smaller injection and loading volume for NH2P-40 3E column.

(1) Injection volume.
To obtain a best resolution, recommended volume is <5 uL. http://shodex.net/index.php?seitenid=1&reiter=r_prod&applic=608 (2) Sample load. A good linear calibration range can be obtained about ~200 ug. Above is just a guideline, thus I would recommend you to optimize the conditions best suit to you.     Additional Comments