Question
measurement of Cacium (20-500 ppm) in the presence of sugar. Currently using Dionex CS12A with suppressor mode. Methan sulfolic acid. Can use YS-50? Also have Hamilton PRP-X100.

 
 

Answer
I do think our best equivalent to Dionex CS12A will be the IC YS-50.
I forgot to ask you which column size you are using, but we do not have an exact equivalent.
If you are using either 2×250 or 4×250 mm columns, I think they are packed with 8 um particle gels while 3×150 is 5 um.
YS-50 is packed with 5 um. So even though this is 4.6 mm i.d., column, I don’t think it will loose too much sharpness compared to 4×250 mm.
If compared to 3×150, there is a potential that YS-50 may show lower resolution.

Attached is the operation manual for YS-50.
http://shodex.net/data/downloads/IC_YS-50_EJ_0512191.pdf
This can be used with suppressor, but with the concentraion of salt that we recommend, it is not necessary.

 
 