Question
using KC-811 for organic acid analysis, 0.1M phosphoric acid. observing unknown peak in the middle, between target acids. wants to find out a way to understand this unknown peak.
wanted to know the nature of KC-811 column and the mobile phase choice.

 
 

Answer
Up to 20% organic solvent can be actually added.
Please see page 8 on the manual and below applications.
http://shodex.net/data/downloads/RSpak_KC-811_EJ_No.506-2.pdf
http://shodex.net/index.php?seitenid=1&applic=838
http://shodex.net/index.php?seitenid=1&applic=839

Or change of acid concentration can change the elution pattern.
http://shodex.net/index.php?seitenid=1&applic=835

Application I mentioned for the analysis of alcohols using water as a mobile.
http://shodex.net/index.php?seitenid=1&applic=816

 
 