Record: 110


Category: column cleaning
Product: KW
Sample:


Question
We have a Shodex KW-6 E806035. The column is used for
analytical purposes just to see how the protein looks in regards to agg
and size. Other people use the column and use reagents with SDS and
NaOAc. Some proteins appear to be degraded on this multi-user column
whereas in another single user set-up (one person knows everything that
is run because he is the one running the column so no degrading buffers)
, it is fine. I was wondering what kind of cleaning protocol you would
reccomend to run before running stds and protein to make sure all
previous reagents and aggregates and proteins are removed. Please
specify reagent % content and time, flow to be
run

 
 

Answer
Thank you for contacting us.
I believe you are using one of our KW series column, however I’m afraid we do not have product named KW-6.
Would you please double check the name and if possible where you purchased the column from?
Do you see any back-pressure change when you suspect the protein adsorption?

As a general information, to clean KW columns series, you may use following steps.

When observing the high pressure, reverse the flow and introduce the eluent at slower flow rate (~1/3 of regular flow rate).

When suspecting protein adsorption, use ~0.3M NaCl solution to clean the column.
*do not leave NaCl solution in the column for a long time period since it may cause rust to the column housing.
Use x10 column volume of the cleaning solution.

When suspecting organic substance adsorption, clean with DI water first, and then clean with organic solvent such as MeOH.
Storing the column with 20% MeOH is recommended over sodium azide solution, since NaN3 is a basic eluent and KW series is not recommended to use under basic conditions.

 
 