Shodex protein analysis columns

Record: 513


Category: Column Selection
Product: KW/DEAE
Sample:


Question
separation of protein from human serum. Range in 10K-500K Da, they also have different pI. Has used DEAE column but not SEC. in the future interested in collecting (prep scale up to ~ 10 mg)

 
 

Answer
Recommended loading volume for KW-800 series column is <0.5 mg (BSA) to obtain a good resolution. http://shodex.net/index.php?seitenid=1&reiter=r_results&applic=561 While for DEAE-825, the test for ovalbumin showed the maximum loading volume is ~1 mg. http://shodex.net/index.php?seitenid=1&reiter=r_results&applic=766 Below is our technical notebook for protein analysis that you might be interested in: http://shodex.net/data/downloads/Shodex_Technical_Notebook_7_Proteins.pdf On page 5, it shows separation differences between different pore-size KW columns. Product information: KW-803 http://shodex.net/index.php?seitenid=1&reiter=r_results&prodid=765 KW-G (guard column) http://shodex.net/index.php?seitenid=1&reiter=r_results&prodid=1217 KW-2005 (preparative size column) http://shodex.net/index.php?seitenid=1&reiter=r_results&prodid=822 DEAE-825 http://shodex.net/index.php?seitenid=1&reiter=r_results&prodid=43     Additional Comments

image_pdfDownload or Print PDF