Shodex SEC columns

Record: 294


Category: Column Selection
Product: KW,SB
Sample: protein with lipid anckers


Question
trying to separate proteins (8,15, 50, 80K Da etc) which have lipid ends. Currently using C4 column, separation is good but low recovery. Heard about Shodex columns and interested in.

 
 

Answer
As I explained over the phone yesterday, there are two options under our GFC (SEC column for aqueous mobile) columns.
PROTEIN KW series is our silica-based SEC column, suitable for protein separation (as you can tell from the name). The eluent selection includes buffers such as phosphate, tris hydrochloric, acetate buffers (pH between 3-7.5). Also to prevent proteins from ionic-interaction with the packed gel, salts are added to the column: Typicall used are 0.1-0.3M of Na2SO4, K2SO4, or (NH4)2SO4. Or as you asked me, surfactant such as SDS and Brij can also be used, as well as 100% MeOH and ACN. Please see page 44 on the catalog linked below to see its specifications.
http://shodex.net/data/downloads/Shodex_2011-2012_Catalog.pdf

Also I am attaching the operation manual for your information.
http://shodex.net/data/downloads/PROTEIN_KW-800series_EJ_No.615-5.pdf

For the separation of proteins with MW ~8,~15, ~50 & ~80KDa, the pore size you would be looking at is KW-802.5, 400A (maximum pore size) which has exclusion limt of 150KDa. Please see below application which may provide you some idea of separation you will see with KW-802.5.
http://shodex.net/index.php?seitenid=1&applic=565

Compared to KW series, SB column (pg 46 on the catalog) is our polymer-based GFC column. SB series can be used in wider mobile phase conditions: pH range 3-10, 100% MeOH, 75% ACN, and 100% DMF. However, compared to silica-based KW column, it has less theoritical number i.e., the separation is not going to be as good as that of KW series. If you do not have pH restriction, I would recommend KW over SB. If you do, SB-804 HQ will be the size you may want to.
Operation manufal SB: http://shodex.net/data/downloads/OHpak_SB-800HQ,SB-2000_series_EJ_No,1141-3.pdf

But again as I mentioned yesterday, that we do not know the reason for low recovery of your samples when analyzed by C4 column.
If the cause is due to the silica/residual silanol on C4 packing gel, there is a potential that SB column may provide better recovery rate.
I’m sorry that I cannot provide you a definite recommendation but I hope the information helps..

 
 