Skip to main content
Print

Starch analysis

Record: 177


Category: Specifications
Product: SB & KW
Sample: protein, starch


Question
Starch analysis

 
 

Answer
1. For the analysis of starch, we do recommend our OHpak SB series columns.
Below is an application using SB-806 with DMSO/DMF mobile phase.
http://shodex.net/index.php?applic=139

The mentioned TSK columns (TSK G4000PWxl and 5000PWxl) our equivalent will be SB-803 and SB-804 respectively.
As Ritsuko mentioned, the exclusion limits for SB series are:

SB-803 HQ: 100, 000 (pullulan)
SB-804 HQ: 1,000,000
SB-806 HQ: 20,000,000

If you are unsure which one to choose from, please let me know the MW (biggest & average) of your target compound.

While our equivalent to PL gel columns are GPC (organic solvent type, gel permeation chromatography) columns.
The below example was obtained by KD-806M, one of our GPC columns.
http://shodex.net/index.php?seitenid=1&applic=1293
(The “M” indicates mixed-gel type, and contains gels with different pore sizes. The exclusion limit is also 20,000,000.)

Please see details of column specifications in the linked catalog: pg46 (SB) and pg54 (KD).
http://shodex.net/data/downloads/Shodex_2011-2012_Catalog.pdf

2. Protein analysis
We recommend our KW series (pg.44 on the catalog) for the analysis of protein.

One of your reference indicated TSK G4000 and G3000swxl, our equivalent will be KW-804 and KW-803.
e.g.http://shodex.net/index.php?seitenid=1&applic=567

 
 
Additional Comments