Skip to main content
Print

Sugar separation

Record: 320


Category: Column Selection
Product: SUGAR/NH2P
Sample: Stachyose, Raffinose, Sucrose, Glucose, Galactose and Fructose.


Question
One of my customer would like to separate the following sugars: Stachyose, Raffinose, Sucrose, Glucose, Galactose and Fructose.
I looked in you’re catalogue and in the table I found out that the more suitable column would be the NH2P-50 4E, but there’s still no big retention time difference between glucose and galactose.
Do you think that this column would do the job for those six compounds? Would it work with a refractive index or PDA detector?

Do you have any other column that would be better or do you have any suggestions in term of conditions to propose?

 
 

Answer
Like you said, the separation between glucose and galactose is not so great by NH2P-50 4E, but could be acceptable (depending on the concentration too).
http://shodex.net/index.php?seitenid=1&applic=617
The separation might be better by using a NH2P-40 3E.
Although, I am concerning that the retention time of stachyose is going to be really long (~35min).

Another option will be using either SP0810 or SC1011.
They can provide better separations between glucose and galactose, but then separation between stachyose and raffinose suffers..
http://shodex.net/index.php?seitenid=1&applic=1009
http://shodex.net/index.php?seitenid=1&applic=624

And yes, both method can be used with RI detector.
Would you please show those applications to your customer to see if they can be acceptable?

 
 