Record: 204


Category: Column Selection
Product:
Sample:


Question
1.there is a client who is using the column C1 Alesorb XT 4.6 x 100 mm 5 mm , from Anacrom but they don’t sell this column anymore. They use this column for impurities of clorambucilo . Do we have an equivalent ?

2.The other one is Microna C8 4.6 x 333 mm 3 mm and they use it for analyzing paroxetina.

1. 3.The other is Thermo RP8/cation 4.6 x 250 mm 5 mm, they use this column and only last for 2 analysis, they wash the column with ACN 100%, and then with ACN/H2O and then finally with water but the column is clogged . The application is for -rosiglitazone-metformin

4. The other one is Onmipack PEX-500 4.0 x 250 mm from Dionex for the analisis of Metformin.

 
 

Answer
1. We do not have RP-C1 columns. By looking at the structure of clorambucilo, I think general RP column should work..but there must be a reason (too much retention on C18) that the customer was using a C1 column. If they cannot find an alternative C1 column, which I doubt..they could try Silica 5C4 4D. but also this is 150 mm, so the analysis is going to be even longer.. not my best recommendation.

2. Silica 5C8 4E would be the closest.
The column length (333 mm) is so weird.. but I couldn’t find more information on this column. Also if this is a 300 mm column, seems bit long for 3um gel.. Anyway, we don’t have 3um C8 column nor 300 mm length.

3. Again, I cannot find the information on Thermo RP8/cation. But I’m guessing this is a silica based column. In that case, our polymer-based column can be a good recommendation, since the main target analyte is a basic compound. Although, I’m not really sure what the customer need to separate from what.. The possibility I’m thinking is NN-414, if they need RP/cation type. If they can use general RP, ODP50 might be a choice. – I think Ana should find out more about this inquiry, I think we have a good chance of recommending something suitable with ours.

4. Could this be Dionex OmniPac PCX-500 mixed-mode column?? Also is this for the same customer as for #3 question? NN-414 has similar function. But the PCX-500 says it can be used between pH 0-14, while NN-414 is 2-12. PCX-500 is a PEEK housing column, NN-414 is a stainless housing. So we need to make sure the working condition of NN is okay.

 
 
Additional Comments