Question
PISA is one of the most important companies in Mexico city and they are asking us to find some applications for the analysis of Shampoo ( pets)

The 1st shampoo :
Olein sulfonato s?dico —-Olein Sodium sulfate
Lauramida D.E.A. Lauramide D.E.A.
Glicerina —-Glycerol

The 2nd shampoo :

PYRIDOXINE HYDROCHLORIDE
Sodium salicylate (salicylic acid)
ZINC GLUCONATE
Melaleuca
Piroctone olamine

 
 

Answer
Regarding the 1st shampoo.
I could not find the details of Olein Sodium sulfate.
Is that CH3(CH2)8=(CH2)8SO3Na?
If so, I think Shodex GF-310 column may be usable.
Please refer to the attached file.

(See attached file: GF Palmitate.pdf)

I think the elution of Olein Sodium sulfate will be the former of that of palmitic acid, because this separation mode is SEC.
And although Lauramide is not analyzed, this compound may be also eluted after the elution of Olein Sodium sulfate by the same reason.

Regarding the “2nd shampoo.

We have the separation data for PYRIDOXINE HYDROCHLORIDE and Sodium salicylate.
Please refer to the following page.


http://www.shodex.com/english/dc050211.html


http://www.shodex.com/english/dc040229.html

However, we are sorry but we do not have the application data for ZINC GLUCONATE, Melaleuca, and Piroctone olamine.

Melaleuca is a kind of oil. So, the organic SEC may be needed.

I think it needs to use SEC or HILIC for the analysis of ZINC GLUCONATE as the following analysis, for example.

http://www.shodex.com/english/dc061334.html

http://www.shodex.com/english/dc030402.html

Piroctone olamine is a kind of basic compound. So, I think reversed phase analysis as the following page may be usable.
http://www.shodex.com/english/dc040231.html

The 2nd shampoo is including higher molecular weight compounds like guar gum or PEG.
It may be needed to remove these before the analysis of the target.
And the low molecular weight compounds like colorants or conservatives may interfere the separation of the target. In also that case, it needs to remove them from the sample solution.

 
 