Technical question Latin America Colombia Warfina, Barbital ,4hidroxinoneal)

Record: 181


Category: Column Selection
Product: RP?
Sample: Warfina, Barbital ,4hidroxinoneal


Question

 
 

Answer
I don’t think the customer is separating three target compounds from each other…
Are they looking for a column suitable for all of the three analyses?

Especially, warfarin analysis, I think isomer determination is important…
http://shodex.net/index.php?seitenid=1&applic=450

We have an application for barbital using ODP2. For the basic compound, ODP2 has the strength. If the customer can use column switching method (pg 6 on the technical notebook #6), that will be even better.
http://shodex.net/index.php?seitenid=1&applic=1217

I think regular C18 can be used for 4-hydroxynonenal analysis.
As you said, C18 should be able to separate all three by looking at their structures.. but ODP probably give a better performance.
Also another thing should be considered is the detector.
What type of detector will be used?
Shodex C18 are only available in 4.6 mm i.d. If the customer is using MS/MS, 2.0 mm i.d. will be better, so ODP has more option.

 
 