Technical question ( phenol resins)

Record: 194


Category: Column Selection
Product: KF
Sample:


Question
They have a client who is working with phenol resins and is using our KF-801 y KF 802.5 columns. He said this column didn’t work properly and then he went to the website and saw the LF 804 column could work better for him. He is asking to please give him an advice.

They do the calibration curve to determine the molecular weight , they handle a wide range of pure phenol resins and they need to cover all the range possible for these resins . Also , They have resins ( resol type )mw= 500 , and novolac of mw = 8000 -12000.
Anna suggested to use KF-400HQ column or KF-804L .They use polyester resins and alkyd resins .

 
 

Answer
We need to know why he is saying that the column didn’t work properly. – does it mean it did not provide good separation of his analyte? Was the analysis time too long?
The exclusion limit of KF-802.5 is 20,000 thus, if his analyte is 12,000 (is this average or the maximum?) KF-802.5+KF-801 should be able to cover.

KF-400 series is the high resolution type, but requires a semi-micro type HPLC system. Unless the customer has it, it will not be any better than KF-800 series.

Is it possible to ask for a chromatogram?

 
 