Product: RI-71 Detector
Sample:


Question
can use toluene?

 
 

Answer
There will not be a problem using toluene as an eluent.

However, when you replace solvent, please take an extra care to wash detector as well as your LC system thoroughly.
If this is not done properly, you will likely observe noisy baseline.

As a guidance:
-To replace an eluent with another eluent that is insoluble to the previously used one, withdraw the old eluent from the line, then wash the line with a solvent which is soluble in both eluents. Then fill the line with the new eluent. For example, if water is replaced with toluene, wash the line with acetone before filing the line with toluene.

-When replacing an eluent containing salt, wash the line with pure water and then acetone before filling the line with the organic solvent. In contrast, when replacing an eluent containing organic solvent with an eluent containing salt, wash the line with acetone first and then pure water before filling the line with the salt-containing-eluent.

 
 
Additional Comments

image_pdfDownload or Print PDF