Question
I have been shopping around for PC and HPLC columns, and I have considered Shodex because of some nice features I have seen from you application notes:

1) High plate numbers
2) Tolerant to wide pH variation
3) Tolerant to solvent chanes

Regarding my applications:

1) HPLC: analysis of residual monomers (free alkylphenols, aliphatic amines), long alkyl aliphatic amines. Secondly, I need to consider other applications such as: reaction (polymerization and oligomerization) monitoring and kinetics (reverse and normal phase)

2) GPC: low molecular weight novolac, phenolic resins, polyphenols, epoxy resins, amine oligomer hardners, REACH compliance, ability to change solvents: THF, CHCl3, DMF

For now, my biggest need is a C18 column to analyze a mixture of long alkyl phenol, and fatty amines. The pH of this material is high, and I may need to use high pH buffers. What do you suggest ? Do you have a specific application note for similar type of analytes ?

 
 

Answer
C18 reversed phase column.
I do recommend our polymer-based columns for high pH analytical conditions.
Most of our polymer-based RP columns can be used in the pH range between 2 to 12 or 13.

Generally, we recommend ODP-50 series for the analysis of basic compounds.
Unfortunately, we do not have a direct application for your analysis, but followings may give us some ides.

Separations of aromatic ketones in high pH mobile
http://shodex.net/index.php?seitenid=1&applic=890

Separation of Catecholamins and their metaboliets.
http://shodex.net/index.php?seitenid=1&applic=420

Separation of fat-soluble vitamins
http://shodex.net/index.php?seitenid=1&applic=688

Our DE-413 column is another polymer-based RP column. But polymer used (polymethacrylate) is different from ODP-50 (polyvinyl alcohol) and the surface is not modified with C18.
Because of the polymer’s unique hydrophilic/phobic feature, it works as a RP column. This would be another option.
Separation of phenols
http://shodex.net/index.php?seitenid=1&applic=885

We also have an application for oligomerization monitoring of polyethylene glycol using this column.
http://shodex.net/index.php?seitenid=1&applic=1048

-GPC column series is one of our strengths and we have many applications.
For an example: http://shodex.net/index.php?seitenid=1&applic=554

Most GPC series (KF/K/KD) can be used with different eluents, but we do recommend to have one column dedicated to one solvent for longer column life.
Gels used for GPC columns are not so rugged for the frequent eluent changes and the eluent replacement does take time.
Since if this is done carelessly, it can cause shrinking/swelling of packed gels.

Our GF series is so called multi-solvent type and has better solvent resistance compared to our regular GPC series.
http://shodex.net/index.php?seitenid=1&applic=1143

 
 