There is a customer looking for a column to analyze the following :

Paracetamol, bromexina, dioxiciclina, piroxicam y guayacol

I can’t find applications on the Shodex website but I noticed in some of these components people use GC , I think Shodex C18 could work but also the ODP , what do you think >?

 
 

Answer
This is for the separation of drugs (some amino compounds), thus regular reversed phase should work.
We would recommend C18.
One example is for the separation of drugs includes paracetamol (acetaminophen) and guaiacol-sulfonate, using ion-pair method.
http://shodex.net/index.php?seitenid=1&applic=1388

By using polymer-based column, ion-pair reagent may not be required.
e.g. analysis of Guaiacol by DE-413L
http://shodex.net/index.php?seitenid=1&applic=830

If the sample contains protein, ODP2 column might be another option.
Paracetamol (acetaminophen) ODP2
http://shodex.net/index.php?seitenid=1&applic=1358

 
 