Question
we do think we’ll be using aqueous phase, since we try to design the peptides we implement to be water-soluble. The samples we will be looking at will be anywhere from <500 to 10,000. In our experiments we're currently binding two PEG-SVA 3400 chains to a peptide to make a single 7000-9000 MW PEG-DA chain, depending on peptide length. We might also do some single bound experiments, with one PEG chain attached to a peptide. Along with the list of columns, could you tell me how long the approximate residence time is for such molecules, and the residence difference between approximately 1000, 3400, 5000, and 8000? Thank for your help, and I look forward to your response.     Answer
In your case, I would recommend you to consider KW-802.5.
http://shodex.net/?seitenid=1&prodid=764
http://shodex.net/data/downloads/Shodex_Technical_Notebook_7_Proteins.pdf

Please see below application. It does not include sample with 5K nor 8K, but you may have a sense of separation between 1K and 3.4K PEG.
The analysis takes about 20 min with flow rate using 0.5 mL/min. You may increase the flow rate ~1.0mL/min to shorten the analysis time, though it may lose some resolution.
http://shodex.net/index.php?seitenid=1&applic=1075

Below shows calibration curve for PEG standards using KW-802.5.
At this condition (1.0 mL/min), you can see that 1K PEG elutes out around 10 min.
http://shodex.net/index.php?seitenid=1&applic=346
As long as your PEG standards have small MW distribution, you should be able to obtain a good separation of each target compounds.
i.e., from this calibration curve, estimated elution time for each 1K, 3.4K, 5K, 8K Da would be 10, 9.5, 8.5, 7.8 ml (min) respectively.
For your information, the separation efficiency of SEC columns can be improved by connecting multiple columns in the series. So if resolution of one KW-802.5 is not good enough, you may use two KW-802.5 columns in series for a better separation.

 
 